* remove hardcoded MODEL_TS80
* #define MODEL_TS100 in sw4stm32/atollic config
* whitespace fix
* use sum of 'defined' instead of logic
* only print handle temp, not PCBVersion
* whitespace fix
* Estimated pinout into the ioc file
* Fix Atollic paths to be somewhat more portable
* Add make command
* Add rough calls to ADC2 [untested]
* Using dual ADC injected modes
* Start both ADCs
* Move some IRQ's to ram exec
* Stabilize PID a bit more
* Add in ideas for tip type selection
* Update peripheral setup to support TS80
* Add tiptype formula / settings struct
* Add function ids to the settings menu
* Rough tip selection
* Rough out new cal routine for simple tips
* Hardware test is fairly close for first pass
* Add Simple calibration case [UNTESTED]
This adds the calibration option that uses boiling water to the calibration menu.
This is untested, and may need gain adjustments before use.
* [Feat] Add some QC testing code
* Typo fix
* Add double button press handler for different rising times
* Add hook for jump to sleep mode
* QC for 9V Works!
* Rough out QC handler, trim out old menu help text thats useless
* QC 9V working... Static all the things (Low on ROM)!
* Static all I2C to save space
* Move QC negotiation into background task so it doesnt block the UI
* Input V display works, tune ADC
* QC 3 steps working
* Start tip R measurements
* Impliment tip resistance
* Fix up the accel position, link in auto QC stages
* Fix tip title
* Tip type settings, Static OLED
* Revert I2C callbacks
* Misc Cleanup
* Better Gain value, need to investiate offset
* Add model warning
* Add TS80 Boot Logo (#367)
* Add TS80 Boot Logo
* Refined
* Moved down by 1px
* Add in power selection 18/24W
* Clean up accelerometer, fix TS100 builds, Fix voltage div cal
* Add rough calls to ADC2 [untested]
* Using dual ADC injected modes
* Start both ADCs
* Move some IRQ's to ram exec
* Stabilize PID a bit more
* Add in ideas for tip type selection
* Add tiptype formula / settings struct
* Add function ids to the settings menu
* Rough tip selection
* Rough out new cal routine for simple tips
* Hardware test is fairly close for first pass
* Add Simple calibration case [UNTESTED]
This adds the calibration option that uses boiling water to the calibration menu.
This is untested, and may need gain adjustments before use.
* Simple Cal Roughly working
* Rough out advanced cal
The flickering on the LCD screen was caused by the OLED DMA taking slightly longer than the delays, so the tail end would flicker if the buffer was cleared before it finished writing.
In future may want to double buffer the LCD.
Fixes#290
This shows the tip temp on the simple home screen if the temp is > 50C.
This removes the tip temp warning for all users, as now everyone can see the tip temp.
#187